Canonical Question

Resp – Oximetry

V5 C9.i Historical V4 F12.i 5 appearances

Master answer

Principles

Beer-Lambert Law

Hb absorbance at 660nm and 940nm utilized

\[ R \; = \; { {Pulsatile_{660} \; / \; Non-Pulsatile_{660} } \over { Pulsatile_{940} \; / \; Non-Pulsatile_{940} } } \]

using healthy volunteers derived R values
Sats 100% R = 0.4
Sats 85% R = 1.0
Sats 50% R = 2
Sats 0% R = 3.4

Image Source: https://commons.wikimedia.org/wiki/File:Oxy_and_Deoxy_Hemoglobin_Near-Infrared_absorption_spectra.png

Pulse oximeter

Limitations

PATIENT FACTORS
Low or High SpO2Low SpO2Normal or High SpO2
Met-Hb
Sulph-Hb
Poor perfusion of finger
Movement artifact
Venous pulsations
Fingernail polish
Intravenous pigmented dyes
Haemoglobinopathy
Anaemia with co-existing hypoxia
Carbon Monoxide poisoning
EQUIPMENT FACTORS
– Ambient light interference
– Poorly fitting probe
– Assay calibrated using healthy volunteers only down to SpO2 80%. Unknown significance if tested SpO2 less than 80%
PHYSIOLOGICAL FACTORS
– Due to O2 dissociation curve, insensitive to changes above PaO2 80mmHg
– Does not measure tissue oxygenation

Absorption spectra confounded by:

Source: https://pubs.asahq.org/anesthesiology/article/105/5/892/6808/Measurement-of-Carboxyhemoglobin-and-Methemoglobin

Sakurai 2016

Exam appearances

ExamExact wordingRelationshipSuccess
2008A Q03 Describe the principles of measurement of arterial haemoglobin oxygen saturation using a pulse oximeter. Outline the limitations of this technique. historical_member
2025A Q18 (a) Describe how arterial haemoglobin oxygen saturation is measured using a pulse oximeter including the underlying scientific principles (60% of marks). (b) Outline the limitations and sources of error of this method (40% of marks). safe_repeat 76.00%
2013B Q05 Describe the principles of measurement of arterial haemoglobin O2 saturation using a pulse oximeter. (60% of marks) Outline the limitations of this technique. (40% of marks) historical_member
2014A Q07 Outline the principles underlying pulse oximetry. ( 80% of marks) Briefly describe the effect of an elevated level of the following upon pulse oximetry values. (20% of marks) a. Carboxyhaemoglobin b. Methaemoglobin historical_member
2021A Q17 Describe the principles of measurement of arterial haemoglobin O2 saturation using a pulse oximeter (60% marks). Outline the limitations of this technique (40% marks). historical_member